The hawkone was returned connected to a cutter driver.
No ancillary devices were included.
The hawkone was visually inspected and it was noted that the guidewire tubing was damaged.
The guidewire lumen of the rotating tip showed that the proximal end was buckled and pointed upwards from the tip.
A yellow/green debris was observed between the rotating tip lumen and the housing lumen.
The debris was likely ptfe from the capture wire of the spider fx.
The guidewire tubing of the housing segment showed a zipper tear throughout the segment.
It was discovered the yellow/green debris was observed on the outside of the torque shaft adapter.
The drive shaft was bent at an approximate 90 degree bend at the strain relief.
No other damages to the hawkone was observed.
Functional testing was performed and a 0.
014" guidewire from the lab was front-loaded.
The guidewire entered the tip, but was protruding out from the beginning of the tear.
The guidewire was back-loaded and the tear was evident throughout the segment.
The guidewire could not be loaded.
If information is provided in the future, a supplemental report will be issued.
|