• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION S7 SYSTEM; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION S7 SYSTEM; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733856
Device Problem Imprecision (1307)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2018
Event Type  malfunction  
Manufacturer Narrative
A manufacturer representative went to the site to test the equipment.During testing the reported issue was unable to be replicated.The system passed the system checkout and was found to be fully functional.No parts were replaced.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation device being used for a transsphenoidal procedure.It was reported that during the case, the surgeons alleged that there was an inaccuracy of about 2 millimeters.The site registered the patient, draped the patient, had the ear, nose & throat (ent) surgeon open a path in the sinuses, and then the neurosurgeon came in to resect the tumor.The tumor in this case was a very small pituitary tumor.The ent surgeon did not actively use navigation but did check accuracy during the procedure and noticed about a 2-millimeter inaccuracy from the beginning of the case.This was also observed by the neurosurgeon.The site proceeded with the case by accounting for the inaccuracy.There was no surgical delay due to this issue, and there was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8136225
MDR Text Key129461062
Report Number1723170-2018-06084
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2013
Device Model Number9733856
Device Catalogue Number9733856
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age57 YR
Patient Weight64
-
-