PERFUSION SYSTEMS Y ADAPTER; APPARATUS, AUTOTRANSFUSION Back to Search Results
Model Number BTC93
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/21/2018
Event Type  malfunction  
Manufacturer Narrative
Product analysis: visual analysis: visual inspection shows evidence of a crack at the base of the "y" connector.Performance analysis: pressure integrity testing was performed at 3 l/pm with 23 psi, (1189 mmhg) of back pressure for 10 minutes.During the pressure integrity testing, there was a small leak observed at the base of the "y" connector.Conclusion: the leakage reported only occurs in those assemblies that used cavity 1.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use when the customer was processing the blood, they noted a crack at the y-connector of the btc93.Customer coated it with bone wax and continued with the case.There was no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Y ADAPTER
Type of Device
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
paula bixby
8200 coral sea street ne
mounds view, MN 55112
7635055378
MDR Report Key8137382
MDR Text Key129614092
Report Number2184009-2018-00038
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K812292
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBTC93
Device Catalogue NumberBTC93
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2184009-11-30-2018-003-R
Patient Sequence Number1
-
-