Product analysis: visual analysis: visual inspection shows evidence of a crack at the base of the "y" connector.Performance analysis: pressure integrity testing was performed at 3 l/pm with 23 psi, (1189 mmhg) of back pressure for 10 minutes.During the pressure integrity testing, there was a small leak observed at the base of the "y" connector.Conclusion: the leakage reported only occurs in those assemblies that used cavity 1.If information is provided in the future, a supplemental report will be issued.
|