• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (PRINGY) VOLBELLA WITH LIDOCAINE; IMPLANT, DERMAL, FOR AESTHETIC USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (PRINGY) VOLBELLA WITH LIDOCAINE; IMPLANT, DERMAL, FOR AESTHETIC USE Back to Search Results
Catalog Number 95834
Device Problems Crack (1135); Gel Leak (1267)
Patient Problem No Patient Involvement (2645)
Event Date 11/08/2018
Event Type  malfunction  
Manufacturer Narrative
Further information regarding event, product, or patient details has been requested.No additional information is available at this time.Device analysis: visual analysis of the device indicates 1.0 ml syringe with 0.1 ml of gel remaining in it received without cap, no needle in an opened tray & pack.Cracks are observed at the 3 mm luer lock level.A review of the device history record has been initiated.If any deviations or non-conformances are found, a supplemental medwatch will be submitted.Device labeling: 5.Precautions ¿ failure to comply with the needle attachment instructions could result in needle disengagement and/or product leakage at the luer-lok® and needle hub connection.
 
Event Description
Healthcare professional reported prior to injection with juvéderm volbella® xc, the "end where you put the needle on cracked and product came out." no injury occurred to the patient or injector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLBELLA WITH LIDOCAINE
Type of Device
IMPLANT, DERMAL, FOR AESTHETIC USE
Manufacturer (Section D)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR  74370
Manufacturer (Section G)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR   74370
Manufacturer Contact
suzanne wojcik
301 w howard lane
suite 100
austin, TX 78753
7372473605
MDR Report Key8137429
MDR Text Key130185325
Report Number3005113652-2018-01708
Device Sequence Number1
Product Code LMH
UDI-Device Identifier10888628032460
UDI-Public10888628032460
Combination Product (y/n)N
PMA/PMN Number
P110033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/02/2020
Device Catalogue Number95834
Device Lot NumberV15LA80161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-