Further information regarding event, product, or patient details has been requested.No additional information is available at this time.Device analysis: visual analysis of the device indicates 1.0 ml syringe with 0.1 ml of gel remaining in it received without cap, no needle in an opened tray & pack.Cracks are observed at the 3 mm luer lock level.A review of the device history record has been initiated.If any deviations or non-conformances are found, a supplemental medwatch will be submitted.Device labeling: 5.Precautions ¿ failure to comply with the needle attachment instructions could result in needle disengagement and/or product leakage at the luer-lok® and needle hub connection.
|