We have received the complaint patch for evaluation and we have confirmed the reported incident with this patch.The patch was observed to have delaminated about 3 cm at one end of the patch.At the same end, we also observed a suture that the surgeon had pierced through the patch.No other defects were noted in this patch.This patch was checked by the operating team before the procedure and was found to be acceptable.During manufacturing, each patch is inspected by our manufacturing as well as by the qc staff for such defect.It is highly unlikely for the inspection team would have missed such a visible defect.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.Further, we have not received any other complaints of a similar nature for devices from this lot.Therefore, we believe that it was an isolated incident.We have also performed a review of complaint history for such incidents for last 10 years for this device.The rate of occurrence for such incident is (b)(4) which is in line with our risk documents.Our ifu properly instructs users not to use cutting point armed sutures or traumatic instruments during the procedure as this may damage the patch.The patch was then removed and replaced with another patch.
|