Model Number G53585
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
510 (k) number; k142688.(b)(4).Lab evaluation: the device related to this occurrence underwent a laboratory evaluation on 21 nov 2018.The needle was found to be broken proximal to the sheath extender.Document review: prior to distribution, all echo-hd-19-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-hd-19-c of lot number c1192809 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1192809.Ifu review: the notes section of the instructions for use, ifu0077-4, which accompanies this device instructs the user to inspect the device prior to use : "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use (ifu0077-4 ).Root cause review: a definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible cause could be attributed to the device being used in a flexed or twisted position during the procedure or tortuous anatomy as indicated in the additional information.Summary: complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
Needle broke off."as per complaint form": after the punction, the needle was outside of the scope, the nurse will put out the needle to check the biopsy, the broken needle came out of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Type of Device
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
heather ryan
o halloran road
national technology park
limerick 
061334440
MDR Report Key8140805
MDR Text Key130654345
Report Number3001845648-2018-00564
Device Sequence Number1
Product Code FCG
UDI-Device Identifier00827002535855
UDI-Public(01)00827002535855(17)190119(10)C1192809
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/19/2019
Device Model NumberG53585
Device Catalogue NumberECHO-HD-19-C
Device Lot NumberC1192809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/06/2018
Event Location Hospital
Date Manufacturer Received11/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-