MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42856
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/12/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the procedure, the valve was programmed to 1.5 and then opened onto the surgical filed.On the sterile field, prior to implantation into the patient, the valve pressure was checked manually with a manometer and found to be at 4 mm h20.It was stated that this does not correlate to a setting of 1.5 and would have led to shunt overdrainage (due to too little resistance).The valve was replaced.
 
Manufacturer Narrative
The returned valve was patent.The valve met the requirements for valve flux, siphon, reflux, and leak testing.The valve did not meet the requirements for pressure-flow and preimplantation testing.Particulates were also observed on the interior and exterior of the valve.It is not known how or when these particulates entered the valve.The instructions for use cautions that, ¿introduction of contaminants could result in improper performance of the shunt system.Particulate matter that enters the shunt system may result in shunt occlusion or may also hold pressure/ flow controlling mechanisms open, resulting in overdrainage.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, SMALL
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8151597
MDR Text Key130060947
Report Number2021898-2018-00558
Device Sequence Number1
Product Code JXG
UDI-Device Identifier00643169975828
UDI-Public00643169975828
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Model Number42856
Device Catalogue Number42856
Device Lot NumberE47049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-