BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115307
Device Problem Misassembled (1398)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/14/2018
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a prismaflex set ckt was misassembled.During continuous renal replacement therapy (crrt) using a prismaflex control unit, it was discovered that pbp line and effluent lines were crossed/misassembled.The misassembled device was discovered during patient treatment.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection on the sample photographs was performed and evidence was shown that the pbp and effluent lines were incorrectly assembled on the support plate.The reported condition was verified.The cause of the reported condition was an assembly issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
na
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8152764
MDR Text Key130661462
Report Number8010182-2018-00167
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K080519
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2020
Device Catalogue Number115307
Device Lot Number18D1203V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-