• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB CUT BLK RM/LL; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB CUT BLK RM/LL; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES Back to Search Results
Catalog Number 202440200
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Instrument has our pin jammed in it, cannot remove it.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> the device was returned and confirmed the failure mode as reported; a pin was stuck in the block.Once the pin was removed, it was observed that the pin appeared bent, which may have contributed to this complaint.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot
=
> null.Device history batch
=
> null.Device history review
=
> null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI TIB CUT BLK RM/LL
Type of Device
KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8160364
MDR Text Key130244412
Report Number1818910-2018-78066
Device Sequence Number1
Product Code HTZ
UDI-Device Identifier10603295082989
UDI-Public10603295082989
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number202440200
Device Lot NumberSO2033628
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2018
Date Manufacturer Received02/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-