Model Number 8800000006
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2018
Event Type  malfunction  
Manufacturer Narrative
No product has been returned for investigation as product remains in-situ.Radiographs provided confirm a lock screw is not in the fully locked down position.No torque value or device usage confirmed.Unknown if the patient conformed to post-operative activity restriction or suffered a fall.No root cause can be determined as lock screw securing is dependent upon proper rod reduction and torque applied which have not been confirmed.No patient injury reported.Labeling review: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation." ".All implants should be used only with the appropriately designated instrument (reference surgical technique)." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." left in-situ.
 
Event Description
On (b)(6) 2017, patient underwent transforaminal lumbar interbody fusion procedure at l4/5 levels with posterior fixation.Follow up radiographs confirmed a set screw back out at right side of l4 level.A revision procedure is being planned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PRECEPT SPINAL SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE INC.
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
george panfili
7475 lusk blvd
san diego, CA 92121
MDR Report Key8166289
MDR Text Key130493096
Report Number2031966-2018-00173
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517344076
UDI-Public00887517344076
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8800000006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-