Catalog Number 8065741080
Device Problem Suction Problem (2170)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that there was no suction.The doctor changed the handpiece; however, this did not resolve the issue.After replacing the cassette, the handpiece did suction and the procedure was completed with no other issues.
 
Manufacturer Narrative
The lot complaint history was reviewed, this is the second complaint for the finish goods lot; however, the first for this issue for this lot.The device history record shows the product was released per specifications.The wet returned sample was visually inspected and no obvious defects were observed that would have caused or contributed to the reported event.A calibrated console was used to sample and the fluidics management system (fms) cassette primed and tuned with the handpiece successfully and could achieve maximum vacuum.No system message was generated, no fluid or air leaks, and no cracks were observed on the connectors that would have contributed to the reported event.The irrigation and aspiration flow rates were measured and found to be within specifications.Fluid flowed from the balanced salt solution (bss) bottle to the irrigation and aspiration manifolds and continuously to the cassette housing.No occlusion or obstruction was observed during inspection and functional testing.The root cause of the customer's complaint could not be established as the returned fms cassette was evaluated and met specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key8170911
MDR Text Key130646470
Report Number1644019-2018-00293
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Catalogue Number8065741080
Device Lot Number2190854H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Date Manufacturer Received02/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
INFINITI SURGICAL PROCEDURE PAK
Patient Age83 YR
-
-