Model Number 9735602
Device Problem Computer Software Problem (1112)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. Testing revealed that the reported issue could not be replicated. The system then passed the system checkout and was found to be fully functional. If information is provided in the future, a supplemental report will be issued.
 
Event Description
On (b)(6) 2018, (b)(6) (rep, hcp); medtronic received information that, while in a functional endoscopic sinus surgery (fess), it was noted that the electromagnetic field was reduced in size for navigation. The surgeon noted there was not substantial metal interference. There was no reported delay to the procedure due to this issue. There was no reported impact on patient outcome. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION COMPACT NAVIGATION SYSTEM
Type of DeviceEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8174046
MDR Text Key130657199
Report Number1723170-2018-06331
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169672956
UDI-Public00643169672956
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 12/18/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-