• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Mechanics Altered (2984)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The reported malfunction was observed and attributed to a faulty fuse on the digital board.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that during a routine shift check by a clinician, the device was unable to switch modes.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8174732
MDR Text Key130678472
Report Number1220908-2018-03708
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946002343
UDI-Public00847946002343
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2011
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-