Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Vena cava wall penetration/perforation has been reported and may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported cold chills, weakness, nausea, lack of energy, aching bones, and dizziness are directly related to the filter and unable to identify a corresponding failure mode at this time.No relevant notes on neither device or lot number.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.The following allegations have been investigated: tilt, vc perforation, cold chills, weakness, nausea, lack of energy, aching bones, dizziness.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown or unavailable.Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(4).
|
Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Patient code: vessels, perforation of (2135)-listed in the ifu, nausea (1970)- not listed in ifu dizziness (2194)- not listed in ifu.Device code: appropriate term/code not available (3191)-device perforation-listed in the ifu, appropriate term/code not available (3191)-tilt- listed in ifu.Name and address for importer site: (b)(4).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Patient allegedly received an implant on (b)(6) 2011 due to a history of pulmonary embolism.Patient alleges tilt, vena cava perforation.Patient further alleges cold chills, weakness, nausea, lack of energy, aching bones when cold, dizziness.
|