• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 1X40G; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 1X40G; BONE CEMENT, ANTIBIOTIC Back to Search Results
Catalog Number 3003940001
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Corrective action has been initiated for the reported issue.The device will not be returned to the manufacturer.Therefore it will not be analyzed.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.However, a corrective action has been initiated to address the reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It has been reported that before the surgery the outer packaging of bone cement was complete, but there was leakage of powder and the inner sterile packaging is opened.
 
Event Description
It has been reported that before the surgery the outer packaging of bone cement was complete, but there was leakage of powder and the inner sterile packaging is opened.
 
Manufacturer Narrative
(b)(4).The following sections were updated: b3-d4-e1-g1-2-g4-h2-h10 the device was not returned to the manufacturer.Therefore it could not be analyzed.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.According to available data, the most probable root cause is due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFOBACIN BONE CEMENT R 1X40G
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
MDR Report Key8177844
MDR Text Key131767283
Report Number3006946279-2018-00421
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
PMA/PMN Number
K150850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date 07/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Catalogue Number3003940001
Device Lot NumberA746BA1504
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-