The investigation determined that lower than expected results were obtained from both a vitros quality control (qc) fluid and a non-vitros qc fluid using vitros chemistry products amon slides on a vitros 5600 integrated system.The investigation concluded that the most likely assignable cause was a calibration issue.The original calibration parameters were reviewed and deemed to be atypical.The calibrator values all appeared similar to each other, but were very high compared to average.After a new calibration event was executed using a fresh set of calibrators and a fresh vitros amon slide cartridge, both from the same lot, results were attained that were within expectations.The reagent performance after the new calibration was comparable to expectations.Therefore, the reagent can be ruled out as a contributing factor to the event.Furthermore, ongoing tracking and trending of complaints has not identified any signals that would suggest there is a systemic issue with vitros amon reagent lot 1018-0250-2979.Performance testing to verify the instrument was not completed.Although the instrument was not verified, there is no indication that the instrument contributed to the event.
|