(b)(4).Note: this report is being filed for an item number that is not sold in the united states, however this item or similar items are used in products that are sold in the united states by b.Braun medical, inc.We received: one used catheter and other components.Visual inspection of the catheter: the catheter is broken off 89mm away from the tip.There was a scratch mark on the broken of cross section.The fault could be reconstructed in a test "example of contact and damage of puncture needle and catheter": the catheter was pulled out after pushing the cannula tip while applying a light force so as to cut into the catheter.The broken catheter was reproduce.Visual inspection of the needle tip: there was slight damage.There were no abnormalities such as bending or damage in other parts.Review manufacturing record: we checked our manufacturing records and any abnormality was not found.No abnormality was found in the catheter acceptance inspection result.Justification: the fault could be reconstructed.According to customer, defect was found after operation.So, there was no damage on catheter before use.This complaint was not confirmed.
|