• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STERILIZER Back to Search Results
Device Problem Fluid Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/25/2018
Event Type  malfunction  
Manufacturer Narrative
A steris technician arrived onsite to inspect the amsco 400 sterilizer following the reported event. The technician found that the root cause of the reported leak to be a broken bolt on a check valve of the water manifold allowing water to leak out onto the floor. The technician made the necessary repairs, confirmed the unit to be operating according to specification, and returned the sterilizer to service. The amsco 400 sterilizer is not under steris service agreement for maintenance activities. The user facility is responsible for all maintenance activities. No additional issues have been reported.
 
Event Description
The user facility reported that water was leaking from their amsco 400 sterilizer. No injuries were reported; however, procedures were postponed due to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAMSCO 400 STERILIZER
Type of DeviceSTERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key8180215
MDR Text Key131012608
Report Number3005899764-2018-00111
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/19/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/25/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-