In the instructions for use (ifu) it is described how the castors of the table can be retracted to lock the table.If the table is locked, this is registered by a pressure switch.In the ifu it is described that a pedal needs to be pushed upwards for approximately 5 seconds to retract the castors and lock the table.During the reproduction, it was found that in case the pedal is pushed up not long enough, the castors are retracted, but some pressure is left in the hydraulic system for extracting the castors.Since the hydraulic system is not completely pressure free, a change of the center of gravity (by positioning the patient) can lead to a higher pressure in the hydraulic system.This causes the switching of the pressure switch.If the switch is changed, a signal is sent to the controller board of the table.In case this signal is received and a micro switch indicating a tilt of the table of more than 2° is activated, the tables movements are locked.This is done to prevent the tilting of a table set on the castors, since this situation would pose a risk for the patient.We assume that a use error has caused the described malfunction.The customer has not actuated the pedal long enough and this has caused some pressure within the mentioned system.The customer reported that table was set into a trendelenburg position and tilted.We assume that this has caused the center of gravity to shift and the pressure within the system to incline.This has caused the pressure switch to be activated.Since the pressure switch was activated and the microswitch for tilt was activated, the tables functions were blocked.Maquet gmbh provides product failure investigation, analysis and resolution for the device described in this report.
|