• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770600Y
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
It was reported the image is blurry and hard to see.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of system had poor image quality, the image was blurry and hard to see was confirmed.The image showed two dark streaks, the cause is an internal failure in the probe causing the image to show the streaks.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to the customer.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
It was reported the image is blurry and hard to see.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8181883
MDR Text Key131004569
Report Number3006260740-2018-03624
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138287
UDI-Public(01)00801741138287
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770600Y
Device Catalogue Number9770600Y
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Date Manufacturer Received01/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-