Catalog Number 71674570
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2018
Event Type  malfunction  
Event Description
It was reported that during surgery device broke while drilling.A back-up device was used to complete the surgery and there were no procedural delays.There was no harm to patient or staff.
 
Manufacturer Narrative
The associated complaint device was returned and evaluated.A visual inspection of the returned intertan starter drill indicated that it fractured about half way the flutes.The fractured portion was not returned.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 7.0MM COMP SCREW STARTER DRILL
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8184289
MDR Text Key131002048
Report Number1020279-2018-02849
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00885556389393
UDI-Public00885556389393
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71674570
Device Lot Number14MNG0018
Date Manufacturer Received11/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-