• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Calibration Problem (2890)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: a field service engineer (fse) followed up with customer over the phone.While troubleshooting with customer, fse noted that customer replaced the column but didn't prime well.Fse instructed customer to run 10 samples.All samples completed without errors.No further action required by field service.The g8 instrument is functioning as expected.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The g8 operator's manual under chapter 1, introduction and applications, states the following: the time from injection of the sample to the time the specific peak elutes off the column is called retention time.The tosoh automated glycohemoglobin analyzer hlc-723g8 software has been written so that each of the expected fractions has a window of acceptable retention times.If the designated peak falls within the expected window, the chromatogram peaks will be properly identified.When a peak elutes at a retention time not within a specified window, an unknown peak (p00) results.If more than one peak elutes at times not specified by the software windows, each is given a sequential p0x title.In order to keep the peaks within their appropriate windows, it may be necessary to change how fast or slow the buffers are moving through the system by changing the pump flow rate.1.4 storage and stability the unopened tskgel g8 variant hsi column should be stored at 4-15°c in a cool location away from direct sunlight.The column is stable until the expiration date printed on the label.Columns are warranted for 2500 injections.In chapter 5: maintenance procedures, section 5.6 provides detail information on how to replace the column.In chapter 6: troubleshooting states the following 201 calib error is generated when the assay results for the calibrators were unsatisfactory.Check the dilution method, and the column and filters.Verify that the reagents have not expired.Do the values for calib-1 and calib-2 in the parameter screen agree with the assigned values (refer to the instructions for use or the label of the calibrator) 211 peak pattern error is generated when peaks were not separated well operator is instructed to check the samples, buffers, and hemolysis & wash solution.6.4 abnormal chromatograms although the percentage of each hemoglobin component may vary slightly from patient to patient, most whole blood samples will contain six fractions: a1a, a1b, f, la1c+, sa1c, and a0.Chromatograms from patients with hemoglobin variants or unknown peaks not recognized by the analyzer are occasionally seen during routine testing.These patterns may indicate interferences or problems with the assay.Therefore, it is important to use caution when troubleshooting.Review all chromatograms to determine whether the results are valid.In most cases, results for the sa1c% are reportable.In some cases, the sa1c% may be invalid depending on the hemoglobinopathy present, the flow rate, and the condition of the column and reagent system.Remember that all abnormal chromatograms are not necessarily the result of abnormalities in the patient sample.Analyzer problems such as a malfunctioning pump or sampling unit, a column that should be replaced or reagents that are incorrectly placed or have been depleted can also cause abnormal chromatograms.In these cases, sequential chromatograms are usually all affected from the point that the problem began.The most probable cause of the reported event due to operator error: inadequate priming after replacing the column.
 
Event Description
A customer reported getting 201 calibration error on the g8 instrument.The customer stated they are running n lot.Customer replaced all the buffers and performed multiple drain flush procedure.Pressure is 8.87 and instrument has a new column, new filter and retention time (rt) was 0.60 minutes (acceptable range is 0.57 to 0.61 minutes).Technical support specialist (tss) instructed the customer to run five whole blood samples.Customer states there are multiple p00 peaks, hb variant, and check peaks flags.All five samples had these errors.Tss will send a replacement column and instructed the customer to perform more drain flushes and run multiple whole blood samples to try to resolve the flags.The customer received the new column, installed it and calibrated; calibration passed but check peaks, multiple p00 peaks, hb variant peaks and peak pattern error 211 persisted on controls and whole blood samples.A field service engineer (fse) was dispatched to address the reported event, which could have resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6509368123
MDR Report Key8187176
MDR Text Key132055669
Report Number8031673-2018-05328
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/20/2018
Distributor Facility Aware Date11/21/2018
Device Age8 YR
Event Location Outpatient Treatment Facility
Date Report to Manufacturer12/20/2018
Date Manufacturer Received11/21/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-