• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 10/25/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed as per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge field service engineer (fse) evaluated the iabp in the getinge japan office.The fse reported that he was able to reproduce the issue and replaced the batteries to address the failure.All calibration, functional and safety tests were performed and passed to factory specifications.The iabp unit was returned to the customer and cleared for clinical service.
 
Event Description
It was reported that during regular inspection of the cardiosave intra-aortic balloon pump (iabp), the customer discovered that the battery of slot 1 became empty during battery mode in 10 minutes after unplugging ac power from the device.The unit then switched to slot 2, but the battery became empty again soon after and the iabp unit shutdown.Both batteries were fully charged during the event occurrence.The customer reported that aside from this event, both charged batteries worked for 3 hours when the regular inspection had been done.There was no patient involvement or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key8187236
MDR Text Key131335541
Report Number2249723-2018-02177
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-XX
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Device AgeYR
Date Manufacturer Received11/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-