MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Computer Operating System Problem (2898); Inadequate User Interface (2958); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/14/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the display screen of the programmer would freeze.It was also reported that the programmer made an audible noise inappropriately, "similar to noise when using stylus on screen but when programming not possible." despite re-calibrating the stylus and changing stylus pens, the programmer was still unreliable.It was noted that it seemed to be a computer/processor problem not associated with the stylus.The programmer has been returned for repair.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: manufacturer¿s analysis was unable to confirm the customer comment that the programmer would freeze, experienced calibration issues, and that it made a noise; the programmer passed incoming functional testing.The mechanical and electrical parts were inspected, and the speaker was replaced as a preventive measure.The hard drive was reconfigured, and the software was reloaded and updated.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8191795
MDR Text Key131298434
Report Number2182208-2018-02324
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581570
UDI-Public00613994581570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Date Manufacturer Received01/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0406-2019
Patient Sequence Number1
-
-