Model Number 2090 |
Device Problems
Computer Operating System Problem (2898); Inadequate User Interface (2958); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the display screen of the programmer would freeze.It was also reported that the programmer made an audible noise inappropriately, "similar to noise when using stylus on screen but when programming not possible." despite re-calibrating the stylus and changing stylus pens, the programmer was still unreliable.It was noted that it seemed to be a computer/processor problem not associated with the stylus.The programmer has been returned for repair.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: manufacturer¿s analysis was unable to confirm the customer comment that the programmer would freeze, experienced calibration issues, and that it made a noise; the programmer passed incoming functional testing.The mechanical and electrical parts were inspected, and the speaker was replaced as a preventive measure.The hard drive was reconfigured, and the software was reloaded and updated.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
|
|