SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE Back to Search Results
Catalog Number 72200873
Device Problem Electrical Shorting (2926)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2012
Event Type  malfunction  
Event Description
It was reported that during a testing, the device presented a short circuit.No case involved.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with alarm sounding and "short circuit detected" error message on both mdu ports.Cause of alarm and error is shorted electronic components on the main pcb.Six transistors were shorted out and a resistor and was burnt.Reason for electronic component failure is unknown but a shorted out mdu is a possibility.Product passed functional testing with a known good main pcb installed.The complaint investigation has concluded the cause of the failure to be defective electronic components on the main digital pcb.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key8193305
MDR Text Key131333003
Report Number1643264-2018-00991
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010607409
UDI-Public03596010607409
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-