Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.Omsc reviewed the manufacturing history of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.[first time; (b)(6) 2018]: all channels: micrococcaceae (18 cfu / endoscope).[second time; (b)(6) 2018]: all channels: staphylococcus coagulase negative (1 cfu / endoscope).[third time; (b)(6) 2018]: all channels: micrococcaceae (2 cfu / endoscope).[fourth time; (b)(6) 2018]: all channels: pseudomonas alcaligenes (7 cfu / endoscope).The testing result does not clear the (b)(6) guideline.The device had been manually reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus (b)(4).(b)(4) sent the subject device to a third party laboratory for an additional microbiological testing.As a result of the testing, no microbe was detected from the sample collected from all the channels of the subject device.The testing result cleared the (b)(4) guideline.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERSCOPE "CYF-5", EUROPEAN VERSION
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key8194526
MDR Text Key132061603
Report Number8010047-2018-02547
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-