MEDTRONIC XOMED INC. XPS® BUR; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 31315066U
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A healthcare professional (hcp) reported via a manufacturer representative that a small fragment of the bur was chipped while in use by the surgeon during a mastoidectomy procedure.The surgeon was able to locate and remove the fragment which was easily found from the surgical site.There was only a minimal delay in the procedure due to changing of bur and removing the fragment.There was no patient impact.
 
Manufacturer Narrative
Analysis found that visually, a chip was taken out of the cutting edge at the tip.The chip was not retuned but would have measured approximately 0.036¿ by 0.012¿.There were multiple indents on the cutting edge of other flutes proximal to the centerline of the cutting tip.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
uriza shums
6743 southpoint drive north
jacksonville, FL 32216
9043328405
MDR Report Key8196028
MDR Text Key131396264
Report Number1045254-2018-00730
Device Sequence Number1
Product Code EQJ
UDI-Device Identifier20885074524805
UDI-Public20885074524805
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/30/2021
Device Model Number31315066U
Device Catalogue Number31315066U
Device Lot Number0213946791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-