MEDTRONIC EUROPE SARL PERCEPTA CRTP; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
|
Back to Search Results |
|
Model Number W1TR04 |
Device Problems
Premature Discharge of Battery (1057); Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that there was questioned whether there was high battery drain due to power on reset (por) and high previous/post outputs.It was noted the device had reached battery depletion prematurely.It was noted there was rising, high and unstable threshold.The device was reprogrammed.The device and leads remain in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a full electrical reset.Analysis of the device memory indicated that the atrial and ventricular rate histogram data were missing/invalid.If information is provided in the future, a supplemental report will be issued.
|
|