SMITH & NEPHEW, INC. MOT DR UNIT, HNDCONT POWERMAX ELITE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 72200872 |
Device Problems
Overheating of Device (1437); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/19/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that before the procedure, the shaver was plugged in and an error message showed up on dii box, then cord felt hot.A new shaver was used and worked correctly with no delay in the case.
|
|
Manufacturer Narrative
|
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of cord overheating could not be reproduced.Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes each in forward, reverse and oscillate directions.Unit passed functional tests on the dyonics power, dii and dii eip test control units with and without footswitch.Current draw was average for this product and overheating of cord did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
|
Event Description
|
It was reported that before the procedure, the shaver was plugged in and an error message showed up on dii box, then cord felt hot.A new shaver was used and worked correctly with no delay in the case.There were no user or patient injuries.
|
|