• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (PRINGY) VOLBELLA WITH LIDOCAINE; IMPLANT, DERMAL, FOR AESTHETIC USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (PRINGY) VOLBELLA WITH LIDOCAINE; IMPLANT, DERMAL, FOR AESTHETIC USE Back to Search Results
Catalog Number 94731RX
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/15/2018
Event Type  malfunction  
Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event Description
Physician reports that upon starting the application of juvéderm® volbella¿ with lidocaine, after "putting the needle, the tip of the syringe was fractured, the thread was normally grasped, application begun and immediately, physician noticed that the plunger was blocked".It was noted that the syringe has more than a half of its content.Fracture of the syringe in the union with the needle (luer lock).The device touched the patient and patient experienced no consequence.Packaged needle used.
 
Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: one empty syringe of 1.0 ml was received with a tray, but without needle and cap.Two cracks were observed at the 3 mm luer lock level.
 
Event Description
It was clarified that "when the physician opened the syringe it was already fractured.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLBELLA WITH LIDOCAINE
Type of Device
IMPLANT, DERMAL, FOR AESTHETIC USE
Manufacturer (Section D)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR  74370
MDR Report Key8199170
MDR Text Key131504777
Report Number3005113652-2018-01846
Device Sequence Number1
Product Code LMH
Combination Product (y/n)N
PMA/PMN Number
P110033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Catalogue Number94731RX
Device Lot NumberV15LA80125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2019
Date Manufacturer Received03/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age38 YR
-
-