SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2018
Event Type  malfunction  
Event Description
It was reported that, during a knee procedure, the motor drive unit hand control overheated.There was a back-up device to complete the procedure as intended, with neither surgical delay nor user/patient injuries stated.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.Mdu failed functional tests with blade stall error and overheating when power cord was bent near strain relief.Power cord assembly grew warm during testing.After troubleshooting, the cause of blade stall and overheating was observed to be a defective power cord assembly.A visual inspection was performed on the power cords external covering and a large kink was observed near strain relief.It was determined that the power cord has shorted or open internal wiring.Motor and hall board were tested and passed functional testing.The complaint investigation has concluded that this unit has succumbed to physical damage to power cord.Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key8199856
MDR Text Key131599597
Report Number1643264-2018-01008
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/24/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown