• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. HANDPIECE MICS; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. HANDPIECE MICS; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Catalog Number 209063
Device Problems Break (1069); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2018
Event Type  malfunction  
Manufacturer Narrative
¿as part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available".
 
Event Description
Center rotary piece came out of the power.Photo attached.Surgical delay 5 minutes.
 
Manufacturer Narrative
Reported event: it was reported that center rotary piece came out of the power.Product evaluation and results: as per attached picture the failure mode was confirmed as the provided picture shows that the center rotary piece disassociated from the handpiece.Product history review: review of the product history records indicate 25 devices were manufactured under lot no k0aws and 22 devices were accepted into final stock on 03/01/2018.Review of qt18 - 03 - 0002 revealed that the non-conformance is not related to the failure alleged in this compliant.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 209063, prodex lot k0aws shows 00 additional complaints related to the failure in this investigation.Conclusions: the alleged failure mode was confirmed.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
 
Event Description
Center rotary piece came out of the power.Surgical delay 5 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE MICS
Type of Device
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8200169
MDR Text Key131594171
Report Number3005985723-2018-00788
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486030193
UDI-Public00848486030193
Combination Product (y/n)N
PMA/PMN Number
K142530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number209063
Device Lot Number42010218/ 4203551
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-