Catalog Number 209063 |
Device Problems
Break (1069); Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/04/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
¿as part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available".
|
|
Event Description
|
Center rotary piece came out of the power.Photo attached.Surgical delay 5 minutes.
|
|
Manufacturer Narrative
|
Reported event: it was reported that center rotary piece came out of the power.Product evaluation and results: as per attached picture the failure mode was confirmed as the provided picture shows that the center rotary piece disassociated from the handpiece.Product history review: review of the product history records indicate 25 devices were manufactured under lot no k0aws and 22 devices were accepted into final stock on 03/01/2018.Review of qt18 - 03 - 0002 revealed that the non-conformance is not related to the failure alleged in this compliant.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 209063, prodex lot k0aws shows 00 additional complaints related to the failure in this investigation.Conclusions: the alleged failure mode was confirmed.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
|
|
Event Description
|
Center rotary piece came out of the power.Surgical delay 5 minutes.
|
|