Manufacturer Narrative
The production device history record (dhr) for this iabp unit is not required to be reviewed per sop 01-061 since the device manufacture date is greater than one year from the event date.The getinge fse evaluated the iabp and opted to replace the power management board to correct the failure.Full calibration, functional and safety tests were performed and passed to factory specifications.The iabp was cleared for use.
 
Event Description
It was reported that the batteries of the cardiosave intra-aortic balloon pump (iabp) did not charge correctly during a functional test performed by a getinge field service engineer (fse).There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key8200323
MDR Text Key131942104
Report Number2249723-2018-02223
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Device Age YR
Date Manufacturer Received12/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-