Catalog Number 283512
Device Problem Electrical Shorting (2926)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly. Udi: (b)(4). Investigation summary: the device was received at the service center and evaluated. The reported problem of the motor being jammed was not confirmed. Though other problems were identified. There was a short circuit in the motor cable so the motor cable was replaced. Resistance values were out of specification so the hand control set was replaced. The device was repaired, tested, and is fully functional. Although we cannot determine a definitive root cause for the reported problem, one possible root cause is the short circuit in the motor cable, which could have affected the function of the motor. A review of the device history record indicated that this lot of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed. At this point in time, no corrective action is required, and no further action is warranted. However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliate in (b)(6) that post-operatively to an unknown procedure that the micro tornado handpiece with hand controls did not turn; and that the motor was jammed. During in-house engineering evaluation, it was determined that there a short circuit in the motor cable on the device. It was not reported if there was a delay in the surgical procedure; however, it was reported that a spare device was available for use to complete the procedure. There was patient involvement reported. There were no injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMICRO TORNADO HP W HANDCONTROL
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8201058
MDR Text Key131604936
Report Number1221934-2018-55857
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 12/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/27/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured10/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-