If additional information should become available, a supplemental medwatch will be submitted accordingly.
Udi: (b)(4).
Investigation summary: the device was received at the service center and evaluated.
The reported problem of the motor being jammed was not confirmed.
Though other problems were identified.
There was a short circuit in the motor cable so the motor cable was replaced.
Resistance values were out of specification so the hand control set was replaced.
The device was repaired, tested, and is fully functional.
Although we cannot determine a definitive root cause for the reported problem, one possible root cause is the short circuit in the motor cable, which could have affected the function of the motor.
A review of the device history record indicated that this lot of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.
At this point in time, no corrective action is required, and no further action is warranted.
However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
It was reported by the affiliate in (b)(6) that post-operatively to an unknown procedure that the micro tornado handpiece with hand controls did not turn; and that the motor was jammed.
During in-house engineering evaluation, it was determined that there a short circuit in the motor cable on the device.
It was not reported if there was a delay in the surgical procedure; however, it was reported that a spare device was available for use to complete the procedure.
There was patient involvement reported.
There were no injuries, medical intervention or prolonged hospitalization.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|