It was reported that prior to the procedure, the physician was unable to adjust the valve.
This was attempted 5-6 times and the valve was never implanted.
A different valve was used to complete the procedure without issue.
It was noted that the manufacturer representative was able to adjust the valve, however the valve was still replaced as the physician had experienced issues.
|
The returned valve was returned unopened in the original sterile packaging.
Laboratory personnel were able to adjust the valve was adjusted to all performance level settings through the packaging.
Therefore, the nature of the complaint could not be replicated by laboratory personnel.
The valve was not tested for valve flux, reflux, siphon, leak, pressure-flow, or preimplantation testing as it was not related to the complaint.
All valves are 100% tested at the time of manufacture.
If information is provided in the future, a supplemental report will be issued.
|