• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 382533
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during the use of the bd insyte¿ autoguard¿ bc shielded iv catheter the catheter did not have blood control after removing the needle.
 
Manufacturer Narrative
Investigation summary: a complaint history check was performed, and this is the 2nd related complaint reported with the defect/condition of leakage beyond the septum (iag bc) with lot #8226673 regarding item #382533.Dhr review was performed on lot number 8226673.The lot number was built / package on afa line 12 from 21aug2018 through 25aug2018.Review of dhr revealed all required challenge samples, set-up and in process testing was performed in accordance with the quality plans.Review disclosed no reject activity findings throughout the build of this lot that would impact the outcome of the quality of the product relevant to the defect stated in the pir.No quality notification were initiated during the build of this lot number.Investigation conclusion: observations and testing could not be performed because units were not received for investigation of this incident.Root cause description: indeterminate ¿ without the actual sample for evaluation and testing there was no physical evidence to confirm or support manufacturing process related issues for the reported defect.Rationale: a formal corrective action will not be initiated at this time.Customer complaint trends are evaluated on a monthly basis.If the trend of a specific type of complaint warrants a formal corrective action, resources will be assigned at that time.Product quality is evaluated during the manufacturing process with prescribed variable and attributes inspections.These inspections are performed by operators and/or process control technicians to ensure any gross process changes are identified.If defects are observed, disposition of the product, root cause and corrective action are applied according to the quality control plan.
 
Event Description
It was reported that during the use of the bd insyte¿ autoguard¿ bc shielded iv catheter the catheter did not have blood control after removing the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key8203540
MDR Text Key131707280
Report Number1710034-2018-00940
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier30382903825333
UDI-Public30382903825333
Combination Product (y/n)N
PMA/PMN Number
K110443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2021
Device Catalogue Number382533
Device Lot Number8226673
Date Manufacturer Received12/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-