Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis investigation confirmed the reported complaint of failed to operate when inserted.For clarification, the electrical continuity test failed.As a result, the energy function of the instrument was not working when activated on the system.No steam generation was observed when the energy pedal was activated while the grips were holding a wet towel.Further testing indicated the conductor wire was most likely broken at the weld location to the electrode.The root cause of the breakage is unclear since the conductor wire cannot break internally due to external mishandling.Based on the information provided, this event is being reported due to the following conclusion: the energy function and weld damage found during failure analysis suggest that sealing issues could occur.Although, no patient harm, adverse outcome or injury was reported, recurrence of the reported failure mode could cause or contribute to an adverse event.
|
It was reported that during a da vinci-assisted total hysterectomy surgical procedure, the vessel sealer extend instrument failed to operate when inserted.After multiple tries, the customer opened a new instrument and inserted it onto the system.The new vessel sealer extend worked correctly.The planned surgical procedure was completed with no patient harm, adverse outcome, or injury reported.
|