• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9438-05
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, a transmitter failed error occured.No additional event or patient information is available.No product or data was provided for evaluation.The complaint confirmation of the reported transmitter failed error could not be determined.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.The device was visually inspected and passed.Voltage testing was performed and failed due to no voltage.No log data available for review.Confirmation of the allegation and root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
ying chen
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key8204898
MDR Text Key131723606
Report Number3004753838-2018-163558
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9438-05
Device Catalogue NumberSTT-GL-004
Device Lot Number5233559
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received08/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-