CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number VPR-GW-FT14
Device Problems Material Separation (1562); Noise, Audible (3273)
Patient Problems Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2018
Event Type  Injury  
Manufacturer Narrative
Analysis of the reported device is in progress. A supplemental report will be submitted when the device analysis is completed. Csi id: (b)(4).
 
Event Description
The viperwire fractured following an unusual audible noise heard during the procedure in the superior femoral artery, which was 40-70% stenosed. The wire fragment was located in the external illiac artery, but attempts to remove the fragment over the next two hours were unsuccessful. A hematoma formed at the access site. Anticoagulants were administered to the patient and the fragment was then surgically removed. The patient was discharged home the following day.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Type of DevicePERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
morgan hill
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key8206073
MDR Text Key131708546
Report Number3004742232-2018-00394
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151260
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/31/2020
Device Model NumberVPR-GW-FT14
Device Catalogue NumberVPR-GW-FT14
Device Lot Number238076
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data