• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) is not required to be reviewed per getinge standard operating procedure since the iabp was manufactured more than a year before the date of event. A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue. The fse observed an artifact on the screen after power up sequence. To fix the issue, the fse cleaned and reseated the color driver cable and that corrected the display failure. The fse then performed all functional and safety checks to meet factory specifications. Unit passed all functional and safety test per factory specifications. The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during a service test performed by the customer, the cs300 intra-aortic balloon pump (iabp) had a monitor related issue and there were lines on the display. There was no patient involvement, and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key8206079
MDR Text Key131773046
Report Number2249723-2018-02240
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/17/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-