AESCULAP AG VENTRICULAR CANNULA DANDY Back to Search Results
Model Number FF124R
Device Problems Material Protrusion/Extrusion (2979); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date of manufacture is 01/2018 (month/year) day of month is unknown.Investigation results: initial reporter indicated six devices of the same lot number exhibited the reported sharp edge at the drill hole/perforation.Six devices were returned for investigation.Three were in an open and used condition, the other three were returned unused in the original package.The three opened devices all have different manufacture dates, which indicates they are from different lots and none of them share the same manufacture date / lot number of the products returned in package (all three packaged are from same lot number).The device related to this medwatch report was visually and microscopically investigated.It was determined that this device displayed visible damage, including perforations and a scratch.Latex examination gloves were used to determine if the perforation has a sharp edge; the glove was not damaged.The lot number of this device was not provided, therefore, review of device history records for this lot number could not be completed.Review of device quality and manufacturing records were checked for the reported lot number (4509784027) and were found to be according to specification valid at the time of production.No similar incidents were reported for this batch number.The damage present on this device may have resulted from improper handling, or may have been caused by another instrument.According to the instructions for use (ifu), the following must be observed: safe handling and preparation: prior to each use, inspect the product for loose, bent, broken, cracked, worn or fractured components.Do not use the product if it is damaged or defective.Set aside the product if it is damaged.Inspection, maintenance and checks: repeat cleaning and disinfection of products that still show impurities or contamination check that the product functions correctly.
 
Event Description
It was reported that a surgeon has 6 ventricular cannulas that have a sharp edge at the drill hole / perforation that could lead to brain tissue injury when the cannula is retracted.There was no patient injury associated with these devices.All are reported as the same lot/batch number.All medwatch reports associated with this report include: 9610612-2018-00611, 9610612-2018-00623, 9610612-2018-00624, 9610612-2018-00625, 9610612-2018-00626.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRICULAR CANNULA DANDY
Type of Device
CANNULA
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key8206773
MDR Text Key132533295
Report Number9610612-2018-00622
Device Sequence Number1
Product Code HCD
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFF124R
Device Catalogue NumberFF124R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Distributor Facility Aware Date12/18/2018
Device Age11 MO
Date Manufacturer Received12/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-