RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Event date: (b)(6) 2018.Date of report: 31dec2018.International udi # (b)(4).A follow-up report will be submitted once the investigation has been completed.
|
|
Event Description
|
The customer reported a "check vent: blower temperature high¿ occurred.There was no patient involvement.The event date was not specified; estimate used.
|
|
Manufacturer Narrative
|
Date rec¿d by mfr : 23jan2019.The manufacturer¿s international service technician could not confirm or duplicate the reported blower temperature issue.However, the device logs revealed the occurrence of error code "1122" (blower temperature high).The manufacturer¿s international service technician replaced the blower to address the reported problem.The unit was checked overall, run in tested, cleaned, and functionally tested and no abnormality was confirmed.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
Date rec'd by mfr: 05/18/2019.Failure analysis on the returned blower assembly shows that the blower assembly passed all testing.A high blower temperature alarm could not be duplicated.There were no faults found with the blower assembly.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|