RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Event date: (b)(6) 2018.Date of report: 31dec2018.International udi # (b)(4).The manufacturer¿s international service technician confirmed the reported issue and replaced the defective touch screen to address the reported problem.The unit successfully passed the required performance verification test.
 
Event Description
The customer reported touchscreen failure.Touchscreen's center part has no response.The device was not in use at the time of the reported event; therefore, there was no patient involvement.The event date was not specified; estimate used.
 
Manufacturer Narrative
Date rec'd by mfr: 11jun2019.The touchscreen assembly was returned for evaluation.A visual inspection of the touchscreen assembly revealed no damage or contamination.Resistance measurements determined the ul_lr and ur_ll resistance and ul_lr/ ur_ll resistance ratio were out of specification.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8207311
MDR Text Key131767522
Report Number2031642-2018-03126
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2019
Was the Report Sent to FDA? No
Date Manufacturer Received12/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-