MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems Use of Device Problem (1670); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. Livanova (b)(4) manufactures the s5 roller pump. The incident occurred in (b)(4). The reported issue was identified during the serial readout investigation of the s5 roller pump. During the evaluation the reported issue could be traced back to the processor board. Therefore livanova (b)(4) requested the board back for further investigation. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) identified during an serial readout investigation that the s5 roller pump did not work properly due to a failure of the can connectivity. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameS5 ROLLER PUMP
Type of DeviceCONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key8212174
MDR Text Key131939097
Report Number9611109-2019-00003
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 01/03/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No