Model Number 106
Device Problem Premature End-of-Life Indicator
Event Date 12/12/2018
Event Type  Malfunction  
Event Description

It was reported by the surgeon that the patient's generator battery depleted faster than expected. The generator's device history records were reviewed. The generator passed all functional specifications prior to release. The trim test tab operation was performed during a time in which model 106 generators were not being laser-routed. An implant card received indicated that the patient's generator was replaced prophylactically. The suspect product has not been received to date. No further information has been received to date.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 106
Type of DeviceGENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key8214907
Report Number1644487-2019-02408
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Type of Report Initial
Report Date 01/03/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received01/03/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date10/20/2017
Device MODEL Number106
Device LOT Number4593
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location Other
Date Manufacturer Received12/12/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/28/2015
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-