Device evaluated by mfr: returned product consisted of a guidezilla guide extension catheter.The hypotube, collar, distal shaft and tip was microscopically and visually inspected.Inspection revealed a partial separation at the collar and tip damage (misshapen).The damage to the tip is likely attributed to being used in a procedure and with a guide wire.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
Reportable based on device analysis completed on 13-dec-2018.It was reported that the guidezilla was broken proximal to the hypotube transition zone near the hub.A guidezilla guide extension catheter was selected for use.During the procedure, it was noted that the catheter was broken proximal to the hypotube transition zone 0.5 cm from the hub.Furthermore, two non bsc stents broke because of the guidezilla.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.However, device analysis revealed a partial separation at the collar.
|