COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN. Back to Search Results
Model Number L1000
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/29/2018
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical inc.Is currently investigating the reported complaint condition.Once the investigation is completed a follow-up report will be filed.(b)(4).
 
Event Description
Per repair authorization form " customer states when step on pedal and only works for second." (b)(4).
 
Event Description
Per repair authorization form " customer states when step on pedal and only works for second." repair order log: 90665.Ref (b)(4).
 
Manufacturer Narrative
Investigation.X- inspect returned samples.Analysis and findings a review of the 2 yr complaint history reveals similar issues.A review of the dhr is not available but not expected to provide relevant information for this complaint.Service and repair confirmed the foot pedal was not properly functioning.The foot pedal is used to activate the device.If not functioning the power to the unit will not flow and it will not cut.The unit is obtained from alsa, an italian manufacturer, and integrated with csi products with minor adjustments that does not involve modifications to the internal board.The foot pedal has been known to fail due to a few reasons.The most common is the diaphragm which acts as the mechanism to make contact (for electrical current) to supply power to the unit.Air displacement pushes on a piston via the diaphragm into a switch to turn the power on.The foot pedal creates the air displacement to the pneumatic switch mechanism located in the housing of the unit.The diaphragm breaks down just enough to lose its seal and cannot function properly.The diaphragm material has been described as a rubber, possibly a latex.Given the appearance of a bad diaphragm it appears to have been exposed to excessive stimuli.Materials like latex are flexible hence the use in this application but its elastic properties can alter over time as well.In this condition, the sealing properties are impacted and it will not function as intended.The root cause for this complaint condition is component related to the diaphragm.Correction and/or corrective action the unit was repaired and returned to the customer.Sustaining engineering has successfully tested a replacement material made of silicone for use in assembly and repairs going forward, eng-test-10341-r.The dfu was also updated to add a safety check via ecn-20444, p/n 35387b.A service bulletin was issued to existing customers informing them to check for this issue and return the unit if needed.All product in fg and sk, as applicable, were reworked to replace the previous versions of the dfus on all applicable products.Correction activity in 2016.Preventative action activity: coopersurgical will continue to monitor this complaint condition for any trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEEP SYSTEM 1000 ESU GEN.
Type of Device
LEEP SYSTEM 1000 ESU GEN.
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull CT 06611
Manufacturer (Section G)
COOPERSURGICAL, INC.
75 corporate drive
trumbull CT 06611
Manufacturer Contact
michael marone
50 corporate drive
trumbull, CT 06611
4752651665
MDR Report Key8223583
MDR Text Key132490802
Report Number1216677-2018-00099
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K952483
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 12/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberL1000
Device Catalogue Number52969
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/23/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-