Model Number W4TR04
Device Problem Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 429878, lead, implanted: (b)(6) 2017. This device was reported as included in the field action. Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action. It is included in the field action in the abundance of caution. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that at a routine follow up, an interrogation showed that a full device reset had occurred several months prior. The device software upgrade was done and the device remains in use. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed. Analysis of the device memory indicated a full electrical reset. Analysis of the device memory indicated that the atrial rate histogram data was missing/invalid. Analysis of the device memory indicated that the ventricular rate histogram data was missing/invalid. This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePERCEPTA QUAD CRTP
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8232438
MDR Text Key132560175
Report Number9614453-2019-00095
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/14/2019
Device Model NumberW4TR04
Device Catalogue NumberW4TR04
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2415-2018

Patient Treatment Data
Date Received: 01/09/2019 Patient Sequence Number: 1
Treatment
407652 LEAD, 407658 LEAD
-
-