Catalog Number 362753 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/20/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that the gel in the bd vacutainer® cpt¿ cell preparation tube with sodium heparin did not separate the red blood cells properly after centrifugation.
|
|
Manufacturer Narrative
|
Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for poor barrier with the incident lot was observed.However, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for poor barrier with the incident lot was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Root cause description: a capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
|
|
Event Description
|
It was reported that the gel in the bd vacutainer® cpt¿ cell preparation tube with sodium heparin did not separate the red blood cells properly after centrifugation.
|
|