BECTON, DICKINSON & CO. BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM HEPARIN; LYMPHOCYTE SEPARATION MEDIUM Back to Search Results
Catalog Number 362753
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/20/2018
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event Description
It was reported that the gel in the bd vacutainer® cpt¿ cell preparation tube with sodium heparin did not separate the red blood cells properly after centrifugation.
 
Manufacturer Narrative
Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for poor barrier with the incident lot was observed.However, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for poor barrier with the incident lot was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Root cause description: a capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Event Description
It was reported that the gel in the bd vacutainer® cpt¿ cell preparation tube with sodium heparin did not separate the red blood cells properly after centrifugation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM HEPARIN
Type of Device
LYMPHOCYTE SEPARATION MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key8237093
MDR Text Key132796555
Report Number1917413-2018-04156
Device Sequence Number1
Product Code JCF
UDI-Device Identifier50382903627535
UDI-Public50382903627535
Combination Product (y/n)N
PMA/PMN Number
K891407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number362753
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received12/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-