Model Number 5392 |
Device Problems
No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Contamination /Decontamination Problem (2895)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator (epg) would not allow adjustment of the settings in the lower screen, such as "the [atrial] sensitivity, upper rate, etc." the issue did not resolve upon restarting the epg.The caller was advised to return the epg for repair.There was no patient involvement.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that the settings of the external pulse generator (epg) would only move up or down by one increment.The epg has been returned for repair.
|
|
Manufacturer Narrative
|
Product event summary: manufacturer¿s analysis was unable to confirm the customer comment that the external pulse generator (epg) would not allow adjustment of the settings; the epg passed all automated and bench testing with no anomalies observed.It was found, however, that the main printed circuit board (pcb) and upper case were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|